ICIMACS for CCDS: Software Notes
8/24/99 : JAM
Disk Synchronization
Users occasionally note that they are having difficulty getting the data disks between
the WC and IC to synchronize on startup (symptom: prospero reports "-Synch" in its
status screen and reports errors after running startup). If full reboots are not solving
the problem, you can attempt to force synchronization of the disks manually:
Check the WC status screen. Near the middle of the screen is a box with disk synchronization
information. If any of the machines' names are followed by the word "-SYNCH", follow the
directions below.
- CB -SYNCH:
- Make sure that caliban is running on the sparc station, then type
>WC REQ INITDISK
in the caliban console window.
- IC -SYNCH:
- On the WC keyboard, type
>IC REQ INITDISK
Note: If the instructions above don't help, then it may be the case that the WC was not able to detect the IC or IR disks
on startup. The only fix is to follow the CCDS startup procedure.
Region-of-Interest
There is at present no region-of-interest selection with the CCDS detector. The
ROI command has no effect on readout with this instrument
Imaging Sciences Laboratory
Revised: August 24, 1999 [jam]